Harmonic B-Mode Imaging
Harmonic B-mode imaging takes advantage of the non-linear oscillation of microbubbles. During harmonic imaging, the sound signal is transmitted at a frequency of around 1.5 to 2.0 MHz and received at twice this frequency. The microbubbles also reflect waves with wavelengths different from the transmitted one, the detectors can be set to receive only the latter ones and create only images of the contrast agent.
Using bandpass filters the transmitted frequency is separated from the received signal to get improved visualization of vessels containing ultrasound contrast agents (USCAs). The signal to noise ratio during the presence of microbubbles in tissue is four- to fivefold higher at the harmonic compared with the basic frequency.
Using harmonic B-mode imaging, harmonic frequencies produced by the ultrasound propagation through tissue have to be taken into account. The tissue reflection produces only a small amount harmonic energy compared to USCAs, but has to be removed by background subtraction for quantitative evaluation of myocardial perfusion.

See also Non-linear Propagation.
Harmonic Imaging
Harmonic imaging relies on detection of harmonics of the transmitted frequency produced by bubble oscillation. This method is widely available on ultrasound scanners and uses the same array transducers as conventional imaging. A major limitation of the use of ultrasound contrast agents is the problem that signals from the microbubbles are mixed with those from tissue. Echoes from solid tissue and red blood cells are suppressed by harmonic imaging.
In harmonic mode, the system transmits at one frequency, but is tuned to receive echoes preferentially at double that frequency, and the second harmonic echoes from the place of the bubble. Typically, the transmit frequency lies between 1.5 and 3 MHz and the receive frequency is selected by means of a bandpass filter whose center frequency lies between 3 and 6 MHz.
Color Doppler and real-time harmonic spectral Doppler modes have also been implemented and show a level of tissue motion suppression not available in conventional modes.

See also Harmonic B-Mode Imaging, and Harmonic Power Doppler.
Non-Linear Imaging
Non-linear imaging is used to detect primary non-linear components of the received echo. Non-linear methods like harmonic imaging and pulse inversion imaging are designed to detect ultrasound contrast agents.

See also Contrast Pulse Sequencing, and Power Modulation.
Power Modulation
Power modulation is a non-linear method, based on a multi-pulse technique where the acoustic amplitude (and hence power) of the transmitted pulses is changed. Full and half amplitudes pulses are used to induce changes in the response of the contrast agent. The received echoes from the emitted half amplitude pulse are adjusted with the full amplitude pulse and this pairs of pulses are subtracted. Power modulation is used to separate contrast agent echoes at low mechanical index, allowing real-time perfusion imaging.
Power modulation can be used with a low frequency wide band transducer to increase the depth and transmit the sound beam homogenous allowing ultraharmonic imaging.
Pulse Inversion Doppler
Selective detection of the microbubble contrast medium can be enhanced by Doppler processing that removes signals with zero Doppler frequency shifts. This will remove tissue harmonics. By detecting overlong bursts of inverted pulses and using Doppler detection methods, very high sensitivity to microbubbles can be achieved. The bubbles can be detected at sufficiently low incident power levels to avoid destroying them. Pulse inversion Doppler has demonstrated the first real-time images of myocardial perfusion using perfluorocarbon gas agents.

See also Pulse Inversion Imaging, Myocardial Contrast Echocardiography, and Perfluorochemicals.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]