Oscillation
Oscillation is a rhythmic periodic motion. Microbubbles, for example oscillate at their natural frequency at which they resonate most strongly.
Harmonic
Harmonic is an oscillation of a system at a frequency that is a simple multiple of its fundamental frequency. The fundamental frequency of a sinusoidal oscillation is called the first harmonic. The second harmonic has a frequency doubled that of the fundamental.

See also Fundamental Imaging, Harmonic Imaging, Subharmonic Imaging and Superharmonic Imaging.
Mechanical Index
(MI) The mechanical index is an estimate of the maximum amplitude of the pressure pulse in tissue. It is an indicator of the likelihood of mechanical bioeffects (streaming and cavitation). The mechanical index of the ultrasound beam is the amount of negative acoustic pressure within a ultrasonic field and is used to modulate the output signature of US contrast agents and to incite different microbubble responses.
The mechanical index is defined as the peak rarefactional pressure (negative pressure) divided by the square root of the ultrasound frequency.
The FDA ultrasound regulations allow a mechanical index of up to 1.9 to be used for all applications except ophthalmic (maximum 0.23). The used range varies from 0.05 to 1.9.
At low acoustic power, the acoustic response is considered as linear. At a low MI (less than 0.2), the microbubbles undergo oscillation with compression and rarefaction that are equal in amplitude and no special contrast enhanced signal is created. Microbubbles act as strong scattering objects due to the difference in impedance between air and liquid, and the acoustic response is optimized at the resonant frequency of a microbubble.
At higher acoustic power (MI between 0.2-0.5), nonlinear oscillation occurs preferentially with the bubbles undergoing rarefaction that is greater than compression. Ultrasound waves are created at harmonics of the delivered frequency. The harmonic response frequencies are different from that of the incident wave (fundamental frequency) with subharmonics (half of the fundamental frequency), harmonics (including the second harmonic response at twice the fundamental frequency), and ultra-harmonics obtained at 1.5 or 2.5 times the fundamental frequency. These contrast enhanced ultrasound signals are microbubble-specific.
At high acoustic power (MI greater than 0.5), microbubble destruction begins with emission of high intensity transient signals very rich in nonlinear components. Intermittent imaging becomes needed to allow the capillaries to be refilled with fresh microbubbles. Microbubble destruction occurs to some degree at all mechanical indices. A mechanical index from 0.8 to 1.9 creates high microbubble destruction. The output signal is unique to the contrast agent.
Ultraharmonic Imaging
Ultraharmonic is an oscillation at a frequency that is a rational multiple of that of its fundamental sinusoidal oscillation, for example 1.5 or 2.5 times the fundamental frequency. Ultraharmonic imaging is a method to eliminate tissue artifacts and therefore increase contrast to tissue ratio.
Also called Superharmonic Imaging.

See also Power Modulation.
Cavitation
Cavitation is any activity of highly compressible transient or stable microbubbles of gas and/or vapour, generated by ultrasonic power in the propagation medium. Cavitation can be described as inertial or non-inertial. Inertial cavitation has the most potential to damage tissue and occurs when a gas-filled cavity grows, during pressure rarefaction of the ultrasound pulse, and contracts, during the compression phase. Collapses of bubbles can generate local high temperatures and pressures. Transient cavitation can cause tissue damage.
The threshold for cavitation is high and does not occur at current levels of diagnostic ultrasound. The introduction of contrast agents leads to the formation of microbubbles that potentially provide gas nuclei for cavitation. The use of contrast agents can lower the threshold at which cavitation occurs.

Types of cavitation:
Acoustic cavitation - sound in liquid can produce bubbles or cavities containing gas or vapour.
Stable cavitation - steady microbubble oscillation due to the passage of a sound wave.
Transient cavitation - short-lived cavitation initiated by the negative pressure of the sound wave.

Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]