Ultrasound Modes
Contrast Modes I
Contrast Imaging Techniques
Many different contrast imaging techniques have been developed. Most are either variations, hybrids, or combinations of the following ultrasound techniques:

See also Coherent Contrast Imaging, Ultrasound Picture and Targeted Contrast Imaging.
• View DATABASE results for 'Contrast Imaging Techniques' (3).Open this link in a new window.
Harmonic Imaging
Harmonic imaging relies on detection of harmonics of the transmitted frequency produced by bubble oscillation. This method is widely available on ultrasound scanners and uses the same array transducers as conventional imaging. A major limitation of the use of ultrasound contrast agents is the problem that signals from the microbubbles are mixed with those from tissue. Echoes from solid tissue and red blood cells are suppressed by harmonic imaging.
In harmonic mode, the system transmits at one frequency, but is tuned to receive echoes preferentially at double that frequency, and the second harmonic echoes from the place of the bubble. Typically, the transmit frequency lies between 1.5 and 3 MHz and the receive frequency is selected by means of a bandpass filter whose center frequency lies between 3 and 6 MHz.
Color Doppler and real-time harmonic spectral Doppler modes have also been implemented and show a level of tissue motion suppression not available in conventional modes.

See also Harmonic B-Mode Imaging, and Harmonic Power Doppler.
• View DATABASE results for 'Harmonic Imaging' (44).Open this link in a new window.
Subharmonic Imaging
Ultrasound waves are created at harmonics of the delivered frequency. Subharmonic imaging uses the harmonic oscillation of a system at a frequency that is a simple fraction of its fundamental frequency. The subharmonic response frequencies has half of the fundamental frequency. The second subharmonic has a half fundamental frequency of one half the frequency, and so on.

See also Harmonic Imaging and Superharmonic Imaging.
• View DATABASE results for 'Subharmonic Imaging' (4).Open this link in a new window.
Superharmonic Imaging
Superharmonic imaging uses higher harmonics like third and fourth harmonic to increase the contrast to tissue ratio compared to a second harmonic imaging mode. Second harmonic imaging is better than fundamental imaging, but has limited capabilities to discriminate between tissue and microbubbles, caused by the non-linear propagation of ultrasound.
• View DATABASE results for 'Superharmonic Imaging' (5).Open this link in a new window.
Ultraharmonic Imaging
Ultraharmonic is an oscillation at a frequency that is a rational multiple of that of its fundamental sinusoidal oscillation, for example 1.5 or 2.5 times the fundamental frequency. Ultraharmonic imaging is a method to eliminate tissue artifacts and therefore increase contrast to tissue ratio.
Also called Superharmonic Imaging.

See also Power Modulation.
• View DATABASE results for 'Ultraharmonic Imaging' (2).Open this link in a new window.
Contrast Harmonic Imaging
(CHI) Contrast harmonic imaging is an ultrasound technique to improve the measurement of blood perfusion or capillary blood flow. Based on the nonlinear properties of contrast agents, CHI transmits at the fundamental frequency but receives at the second harmonic. Contrast enhanced echo signals contain significant energy components at higher harmonics (bubbles acts as harmonic oscillators), while tissue echoes do not. Caused by that contrast signal can be separated from tissue echoes by the characteristic signal.
In combination with the pulse inversion technique, CHI promises very high contrast agent sensitivity with high spatial resolution.

See also Ultrasound Contrast Agent Safety and Hemoglobin.
• View DATABASE results for 'Contrast Harmonic Imaging' (9).Open this link in a new window.
Harmonic Power Doppler
(HPD) Harmonic power Doppler is currently one of the most sensitive techniques for detecting ultrasound contrast agents. HPD works by transmitting multiple pulses toward the object to be imaged and detecting the pulse-to-pulse changes in the received echo signals.
Second harmonic bandbass filtering is applied to the received signals to exploit the non-linear behavior of scattering from bubbles (clutter). Harmonic power Doppler operates best at high output levels because of increased contrast destruction, and pulse amplitudes close to the maximum allowed are used much of the time.
With a high mechanical index, non-linear propagation of the sound will cause significant harmonic components from tissue, and the contrast agent to tissue ratio will decrease.
Also called Harmonic Power Angio. See also Multiple Frame Trigger.
• View DATABASE results for 'Harmonic Power Doppler' (5).Open this link in a new window.
I'm living so far beyond my income that we may almost be said to be living apart.
- e e cummings
 
Share This Page
Facebook
Twitter
LinkedIn

Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]