1
term [
] - 36 definitions [
]
Result Pages :
Array Transducer
An array transducer is composed of multiple piezoelectric crystal elements arranged in an array. Arrays are transducer assemblies with a row of elements, used to focus the beam.

Types of array transducers:
Linear array transducer = the arrays are arranged along a line.
Curvilinear or curved transducer = the arrays are arranged along a convex curve. A curved array is similar to a linear array except that the image created is a sector-type.
Annular array transducer = the arrays are arranged in concentric circles.
Rectangular array transducer = the arrays are arranged in a rectangular pattern.

See also Amplitude Shading, Transducer Types, and Transducer Assembly.
• 
View NEWS results for 'Array Transducer' (1).Open this link in a new window.
Linear Probe
A linear probe contains one or more acoustic linear array transducer elements arranged in a line to send pulses of sound into a material. The linear array gives a large probe surface (footprint) and near field. In Doppler mode, a linear probe operates a subset of its elements as a linear phased array and can thus steer the Doppler beam at a selected angle to the imaging beam. This is a popular configuration for peripheral vascular and perivascular scanning.
Probe
In the field of medical ultrasound imaging, the term 'probe' specifically refers to the ultrasound transducer and represent the handheld device that emits and receives ultrasound waves during an examination.
The probe encompasses various components such as the elements, backing material, electrodes, matching layer, and protective face that are responsible for both emitting and receiving the sound waves. Aperture, known also as the footprint, is the part of the probe that is in contact with the body. When the emitted sound waves encounter body tissues, they generate reflections that are received by the probe, which then generates a corresponding signal. In most cases, the probe emits ultrasound waves for only about 10% of the time and receives them for the remaining 90%.
Probes are available in different shapes and sizes to accommodate various scanning situations. The footprint is linked to the arrangement of the piezoelectric crystals and comes in different shapes and sizes e.g. linear array transducer//convex transducer. The transducer plays a huge role in image quality and is one of the most expensive parts of the ultrasound machine. Mechanical probes steer the ultrasound beam driven by a motor and are capable of producing high-quality images, but they are prone to wear and tear. Mechanical probes have been mostly replaced by electronic multi-element transducers, but mechanical 3D probes still remain for abdominal and Ob-Gyn applications.
In summary, the terms 'ultrasound transducer,' 'probe,' and 'scanhead' are often used interchangeably to refer to the same component of the ultrasound machine. Probes consist of multiple components and are available in different shapes and sizes depending on the sonographer's needs.

See also Handheld Ultrasound, Ultrasound System Performance, Omnidirectional, Probe Cleaning, and Multi-frequency Probe,
Refraction Artifact
Different sound velocities in tissue are causing refraction artifacts. With convex elastomer lens transducers, sound beam refraction at the skin interface can alter the transducer's focusing characteristics and beam profile, cause element to element nonuniformity, and cause phase changes in the acoustic wave. These cumulative refraction induced errors degrade the image quality through distortion and loss of resolution. Because the amount of refraction is proportional to the velocity mismatch, the greater the mismatch, the greater the refraction.
Transducer Assembly
A transducer assembly is the configuration of an ultrasound transducer. Electronic array transducers are composed of multiple crystal elements e.g., linear array and phased array transducer assemblies. A handheld mechanical probe may contain a mechanically driven single element transducer, or a rotating wheel transducer. Static B-scanner (obsolete) had a cylindrical probe with a single disk-shaped piezoelectric crystal.
Some probes are designed to be inserted in body cavities (vagina, rectum, esophagus) so that they can get closer to the organ being examined (uterus, prostate gland, heart); getting closer to the region of interest allow for a more detailed view.

See also Endocavitary Echography, Endoscopic Ultrasound, Vaginal Transducer and Rectal Probe.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]