Bolus Injection
A bolus is a rapid infusion of high dose contrast agent. Dynamic and accumulation phase imaging can be performed after bolus injection. Since the transit time of the bolus is only a short time, images with high frame rate show the wash in and wash out of the contrast material. The injection rate and the total injected volume modifies the bolus peak profile. Substantial changes in the concentrations during signal acquisition induce artifacts. Furthermore, the hemodynamic parameters (cardiac output, blood pressure) influence the bolus profile. However, the characteristics of ultrasound contrast agents are favorable with a continuous perfusion.

See also Negative Bolus.
Bubble Destruction
Bubble destruction describes the microbubble shell rupture by ultrasound pulses. The bubble destruction increases with increasing peak negative pressure and decreasing frequency. The mechanical index is an indicator for the effectiveness of microbubble destruction. Contrast enhanced ultrasound relies on bubble rupture to detect bubbles in small vessels.

See also Negative Bolus.
Fetal Ultrasound
Fetal ultrasound is a safe and non-invasive imaging technique used to visualize and monitor the development of a fetus during pregnancy. It employs high-frequency sound waves to create detailed images of the baby, the placenta, and the uterus. Fetal ultrasound provides valuable information about the baby's growth, organ development, and overall well-being. It is commonly used to determine gestational age, assess fetal anatomy, detect abnormalities, and monitor fetal movements and heart rate. This essential tool enables healthcare professionals to ensure the optimal health of both the mother and the baby throughout the pregnancy.
The FDA (Food and Drug Administration) has established regulations governing ultrasound usage, including specific guidelines for fetal ultrasound examinations. These regulations permit an eight-fold increase in ultrasound intensity for fetal scans. They place considerably responsibility on the user to understand the output measurements, the mechanical index (MI), the thermal index (TI) and to use them in their scanning. The primary safety concern in prenatal diagnostic imaging is temperature rise. It is known that hyperthermia is teratogenic. The efforts of investigators have concentrated on defining the temperature increases and exposure times which may give rise to biological effects and on determining the ultrasound levels which might, in turn, lead to those temperature rises.
In fetal ultrasound, the highest temperature increase would be expected to occur at bone and the thermal index with bone at/near the focus (TIB) would give the 'worst case' conditions. The mechanical index and thermal index must be displayed if the ultrasound system is capable of exceeding an index of 1. The displayed indices are based on the manufacturer's experimental and modeled data. However, an independent study has demonstrated significant discrepancies over declared spatial peak time averaged intensity (I-SPTA) output of up to 400%.

See also ALARA Principle, Pregnancy Ultrasound and Doppler Fluximetry in Pregnancy.
M1091
From POINT Biomedical Corp
M1091 is a nitrogen-filled submicron ultrasound contrast agent with a bilayered shell composed of a biodegradable polymer and a human serum albumin outer coating. This agent under development consists of microbubbles that have a peak volumetric diameter of 0.57 μm, and a mean initial diameter of 800 nm, with more than 93% having a diameter smaller than 1 μm. Manipulations in the manufacturing process can generate structures varying in size and wall thickness.
M1091 and M1134 are two agents made of the same polymer with the same relative wall thickness, differing only in the microbubble diameter.
Mechanical Index
(MI) The mechanical index is an estimate of the maximum amplitude of the pressure pulse in tissue. It is an indicator of the likelihood of mechanical bioeffects (streaming and cavitation). The mechanical index of the ultrasound beam is the amount of negative acoustic pressure within a ultrasonic field and is used to modulate the output signature of US contrast agents and to incite different microbubble responses.
The mechanical index is defined as the peak rarefactional pressure (negative pressure) divided by the square root of the ultrasound frequency.
The FDA ultrasound regulations allow a mechanical index of up to 1.9 to be used for all applications except ophthalmic (maximum 0.23). The used range varies from 0.05 to 1.9.
At low acoustic power, the acoustic response is considered as linear. At a low MI (less than 0.2), the microbubbles undergo oscillation with compression and rarefaction that are equal in amplitude and no special contrast enhanced signal is created. Microbubbles act as strong scattering objects due to the difference in impedance between air and liquid, and the acoustic response is optimized at the resonant frequency of a microbubble.
At higher acoustic power (MI between 0.2-0.5), nonlinear oscillation occurs preferentially with the bubbles undergoing rarefaction that is greater than compression. Ultrasound waves are created at harmonics of the delivered frequency. The harmonic response frequencies are different from that of the incident wave (fundamental frequency) with subharmonics (half of the fundamental frequency), harmonics (including the second harmonic response at twice the fundamental frequency), and ultra-harmonics obtained at 1.5 or 2.5 times the fundamental frequency. These contrast enhanced ultrasound signals are microbubble-specific.
At high acoustic power (MI greater than 0.5), microbubble destruction begins with emission of high intensity transient signals very rich in nonlinear components. Intermittent imaging becomes needed to allow the capillaries to be refilled with fresh microbubbles. Microbubble destruction occurs to some degree at all mechanical indices. A mechanical index from 0.8 to 1.9 creates high microbubble destruction. The output signal is unique to the contrast agent.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]