Medical Ultrasound Imaging
Monday, 13 May 2024
• Welcome to
     Medical-Ultrasound-Imaging.com!
     • Sign in / Create account
 
 'Interference' p2
SEARCH   
A B C D E F G H I J K L M N O P Q R S T U V W Z 
Searchterm 'Interference' found in 10 articles
2
terms [
] - 8 definitions [
]
Result Pages :
Linear Array Transducer
Linear array transducer elements are rectangular and arranged in a line. Linear array probes are described by the radius of width in mm. A linear array transducer can have up to 512 elements spaced over 75-120 mm. The beam produced by such a narrow element will diverge rapidly after the wave travels only a few millimeters. The smaller the face of the transducer, the more divergent is the beam. This would result in poor lateral resolution due to beam divergence and low sensitivity due to the small element size.
In order to overcome this, adjacent elements are pulsed simultaneously (typically 8 to 16; or more in wide-aperture designs). In a subgroup of x elements, the inner elements pulse delayed with respect to the outer elements. The interference of the x small divergent wavelets produces a focused beam. The delay time determines the depth of focus for the transmitted beam and can be changed during scanning.
Linear arrays are usually cheaper than sector scanners but have greater skin contact and therefore make it difficult to reach organs between ribs such as the heart. One-dimensional linear array transducers may have dynamic, electronic focusing providing a narrow ultrasound beam in the image plane. In the z-plane (elevation plane - perpendicular to the image plane) focusing may be provided by an acoustic lens with a fixed focal zone.
Rectangular or matrix transducers with unequal rows of transducer elements are two-dimensional (2D), but they are termed 1.5D, because the number of rows is much less than the number of columns. These transducers provide dynamic, electronic focusing even in the z-plane.

See also Rectangular Array Transducer.
Near Field
The near field (also called Fresnel zone) is the proximal part of an ultrasound beam. The Fresnel zone is adjacent to the transducer surface and has a converging sound beam profile. A narrow beam shape is maintained in the near field owing to constructive and destructive interference patterns of sound wavelets emitted from the transducer crystal.
The length of the near field is equal to
r2/l = d2/4l
where r is the radius, l is the ultrasound wavelength in the medium of propagation and d the diameter of the piezoelectric crystal.

See also Beam Pattern, and Sonographic Features.
Point Scatterer
A point scatterer is a reflector with a diameter much smaller than the ultrasound wavelength. The reflection from blood is a typical example of point scattering. Red blood cells are with 7μm versus 0.44 mm wavelength at 3.5 MHz, smaller than any US wavelength. The individual cells are not only the point scatterers, ultrasound is scattered whenever there is a change in acoustic impedance, and in blood such changes are caused by variable cell concentration. These local fluctuations in cell concentration have a spatial extent that is also much smaller than the ultrasound wavelength, and they therefore act as point scatterers.
A point scatterer gives rise to spherical wavelets spreading out in all directions with the scatterer itself at the center of the sphere. The spherical wavelets from one single point scatterer are much too weak to be detected by the transducer, but constructive interference between numerous wavelets will produce backscattering of higher amplitude echoes with parallel wavefronts, also in the direction of the ultrasound transducer.

See also Rayleigh Scattering.
Rayleigh Scattering
Rayleigh scattering is the backscattering of ultrasound from blood. The echoes detected from blood are created through interference between scattered wavelets from numerous point scatterers. Rayleigh Scatterers are objects whose dimensions are much less than the ultrasound wavelength. Rayleigh scattering increases with frequency raised to the 4th power and provides much of the diagnostic information from ultrasound. Doubling the ultrasonic frequency makes the echoes from blood 16 times as strong. The intensity of the backscattered echoes is proportional to the total number of scatterers, which means that the echo amplitude is proportional to the square root of the total number of scatterers.
At normal blood flow, the number of point scatterers in blood is proportional to the number of red blood cells. When blood flow is turbulent, or accelerating fast (e.g. in a stenosis), the number of inhomogeneities in the red blood cell concentration will increase.

See also Scattered Echo.
Real-Time Transducer
Transducers used for the real-time mode are different than for the A-mode, B-, or M-modes. A linear array transducer with multiple piezoelectric crystal elements that are different arranged and fired, transmits the needed larger sound beam.
A subgroup of x adjacent elements (8-16; or more in wide-aperture designs) is pulsed simultaneously; the inner elements pulse delayed with respect to the outer elements. The interference of the x small divergent wavelets generates a focused beam. The delay time determining the focus depth of a real-time transducer can be changed during imaging.
Similar delay factors applied during the receiving phase, result in a dynamic focusing effect on the return. This forms a single scan line in the real-time image. To produce the following scan line, another group of x elements is selected by shifting one element position along the transducer array from the previous group. This pattern is then repeated for the groups along the array, in a sequential and repetitive way.
Result Pages :
 
Share This Page
Facebook
Twitter
LinkedIn
Look
      Ups
Medical-Ultrasound-Imaging.com
former US-TIP.com
Member of SoftWays' Medical Imaging Group - MR-TIP • Radiology TIP • Medical-Ultrasound-Imaging
Copyright © 2008 - 2024 SoftWays. All rights reserved.
Terms of Use | Privacy Policy | Advertise With Us
 [last update: 2023-11-06 01:42:00]